5700 Dew Point Calibration Systems

Dew point calibration system is designed to verify and calibrate SF6 dew point meter, SF6 humidity sensor. The system is combined by a humidity generator and a dew point mirror, capable of producing dew point temperature from -65°C to +20°C.

The system provides dependable and long-term stable dew point output by approriate proportion of both wet and dry flow and control of moisture. It is integrated with a pressure swing dryer that will generate -65°C or even lower dry air or Nitrogen under general laboratory conditions.

The system features appearance, movable, easy operation, high precision and stability.

 

Key Features

 
  • Highly precision and reliability
  • Integrated and movable
  • Easily on screen operation

 

Specification


Dew point range: -65°C~+20°C
Generator mode: Fully automatic, adjustable to continuous mode
Calibration flow rate: 2~8 NI/min, 10~20 I/min (Option) Uncertainty: ±0.1°C dew point
Set point precision: ±0.5°C dew point
Source gas: Air/N2
Operating temperature: +10°C~+30°C
Traceability: Directly to national standard, NPL, NIST and SCS
Power supply: 220~240V AC, 50Hz
Power consumption: 284W
Dimension:
Rack: 595×774×1820mm
Generator: 219×433×450mm
Reference Hygrometer: 485×147×370mm
Weight:
Generator: 18kg
Reference Hygrometer: 10kg
Whole system: 130kg

 