iSF6-UV100 High Precision SO2 Analyzer

Our products are widely used in Gas Insulation Switch (GIS), industrial process control, instrument testing and other fields.

iSF6-UV100 High precision SO2 analyzer is a gas analyzer based on the advanced non-dispersive UV absorption photometry technology, which can detect the concentration of SO2 in SF6 complex gas mixture. It can stably detect the trace components of SO2 in SF6 without cross interference through a special optical technique. The results show that the system can stably detect very small changes in the gas concentration. The unique way of sampling, the high precision flow auto control and the touch screen operation improved the accuracy and ease of use.

The iSF6-UV100 is configurated with the function of pressure and temperature compensation free of Moisture Cross & SF6 Interference. It can measure the SO2 concentration in GIS with high accuracy and stability, and simultaneously detect the gas pressure in the equipment. The analyzer provides flexible measurement modes, allowing the users to make options from the modes of continuous or automatic measurement. It also has a variety of units switchable.

The analyzer is equipped with full-color wide LCD touch screen, which satisfies the high definition contrast and increases the readability. The function of auto start measurement makes the whole measurement easier and more convenient.

The analyzer adopts AC/DC working mode to solve the problem of on-site power supply. Meanwhile, it realized the portable convenience to the users by the visualized battery indication, the rapid charging technology, and the ultra-long standby scheme.
 

 

Key Features

 
  • Response time T90<10s
  • Non-Moisture Cross Interference
  • Low power consumption
  • Pressure compensation
  • Temperature compensation
  • Ultra-long standby time
  • Large capacity for data storage
  • Automatic flow control
  • User self-calibration
  • Smart power and time management display
  • High precision electronic flowmeter
  • AC/DC working, smart charging protection

 

Specification

Measuring principle: Non-dispersive UV absorption photometry
Measuring range: 0~100 ppm
Accuracy: <±1% F.S.
Repeatability: <±1% F.S.
Resolution: 0.01
Response time: T90<10s
Flow rate: 200~500 ml/min
Operating temperature: 5~45℃
RH: 0%~95%,non-condensing
Power supply: DC/AC, AC 110~220VAC,50/60Hz
Battery: 14.8V 10.2Ah,>8 hours standby time
Dimension: 502mm × 401mm × 188mm
Weight: 10.5 kg
Accessories: DN8,DN20,6m Stainless steel armored PTFE tube

 